Anne J. Kleinnijenhuis*, Jan H. Toersche, Frédérique L. van Holthoon, Richard C. Bas, William D. van Dongen 

Journal of Applied Bioanalysis. Vol.1. No.1. pages 26-34 (2015)

Published 15 January 2015. https://doi.org/10.17145/jab.15.005 | (ISSN 2405-710X)

Download PDF

Correspondence:
*Kleinnijenhuis AJ . TNO Triskelion BV, Utrechtseweg 48, 3704 HE Zeist, The Netherlands. Phone: +31 (0)88 866 5007.  †Present address . : ProQR Therapeutics NV, Darwinweg 24, 2333 CR Leiden, The Netherlands.

Citation:
Kleinnijenhuis AJ, Toersche JH, van Holthoon FL, Bas RC, van Dongen WD. A generic sample preparation approach for LC–MS/MS bioanalysis of therapeutic monoclonal antibodies in serum applied to Infliximab. J Appl Bioanal 1(1), 26-34 (2015).

Open-access and Copyright:
©2015 Kleinnijenhuis AJ et al. This article is an open access article distributed under the terms of the Creative Commons Attribution License (CC-BY) which permits any use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Funding/Manuscript writing assistance:
The authors have no financial support or funding to report and also declare that no writing assistance was utilized in the production of this article.

Competing interest:
The authors have declared that no competing interest exist.

Article history:
Received: 13 December 2014, Revised 12 January 2014, Accepted 13 January 2015

Abstract

In this study, we developed a generic bioanalytical workflow providing sensitive, specific, and accurate absolute quantification of therapeutic monoclonal antibodies in serum. The workflow involves magnetic beads coated with protein A to pull-down therapeutic monoclonal antibodies with affinity for protein A from the biological matrix, followed by tryptic digestion and LC-MS/MS quantification of a unique signature peptide, considering of course the matrix of interest and other present mAbs, if applicable. The feasibility of this approach was demonstrated for Infliximab (trade name Remicade) in rat serum. The assigned signature peptide was monitored in the selected reaction monitoring (SRM) mode. Assay variability was determined to be below 20%, except at the QC low level, which was provided through optimization of the sample preparation and monitoring of the LC-MS/MS using a stable isotope labeled signature peptide as internal standard. The 100 ng/ml lower limit of quantification using only 25 µl sample volume, is generally considered as sufficient for pharmaceutical development purposes for monoclonal antibodies.

Keywords

Bioanalysis, monoclonal antibodies, LC-MS, Infliximab

Introduction

In order to quantify proteins in complex biological matrices such as serum or plasma, LC-MS/MS is finding its way to the bioanalytical community [1-6] as a complementary method to ligand binding assays (LBA). Advantages of LC–MS over LBA are that no highly selective biological reagents have to be developed, which can take up to six months to generate and possibly less complications due to immunogenicity [4,7-10]. In contrast to LBAs, the conventional standard for therapeutic protein bioanalysis, which measures the unbound intact 3D domain of the therapeutic protein binding to the assay’s capture reagent, LC–MS measures a unique part of the primary structure of the protein. We developed a generic workflow consisting of immunoprecipitation, using protein A capture followed by LC–MS/MS quantification of a unique signature peptide after tryptic digestion of the target therapeutic monoclonal antibody (mAb) in biological matrices [11]. To optimize the generic approach the “seven critical factors” approach was followed [5]. Essential to the generic methodology are magnetic beads, functionalized with protein A, a 42 kDa surface protein originally found in the cell wall of the bacterium Staphylococcus aureus which binds immunoglobulins. Protein A can be suitable to purify classes, subclasses and fragments of immunoglobulins as well as immunoglobulin complexes [12] when the immunoglobulin of interest has a high affinity for protein A as compared to other affinity capture materials. As a result of the ability of Protein A to capture immunoglobulins it is suitable to concentrate therapeutic monoclonal antibodies which are usually immunoglobulin based. The digestion is combined with disulfide bond reduction and alkylation. To demonstrate the feasibility of this approach, Infliximab (trade name Remicade), a chimeric monoclonal antibody against tumor necrosis factor alpha (TNF-α) used to treat autoimmune diseases, was used [13].

Materials and methods

Chemicals and reagents

Infliximab was obtained from UMC Utrecht as 10 mg/ml solutions. The concentration was verified using Remicade obtained from a local pharmacy. TPCK- trypsin was from Thermo Scientific (Breda, the Netherlands). An 18-mer synthetic signature peptide corresponding to a tryptic fragment (DILLTQSPAILSVSPGER) of the variable part of the light chain of the therapeutic antibody was synthesized by Sigma-Aldrich (Zwijndrecht, the Netherlands). Its stable isotope labelled analogue (DILLTQSPAILSVSPGER[13C615N4]), applied as internal standard, was also synthesized by Sigma-Aldrich. Rat serum was obtained from Harlan (Horst, the Netherlands). PureProteome protein A magnetic beads were obtained from Millipore (Amsterdam, the Netherlands). The other chemical reagents and solvents used were obtained from Merck (Amsterdam, the Netherlands; acetic acid, hydrochloric acid, sodium chloride), Biosolve (Valkenswaard, the Netherlands; acetonitrile, formic acid, methanol), Fluka (Zwijndrecht, the Netherlands; ammonium bicarbonate, calcium chloride and sodium hydroxide), Sigma-Aldrich (dithiothreitol, glycine, iodoacetamide, monosodium phosphate, trifluoroacetic acid) and Calbiochem (Amsterdam, the Netherlands; tris(hydroxymethyl)aminomethane). 

Preparation of calibration and QC samples

Calibration standards were prepared by serial dilution of Infliximab to 0.05, 0.10, 0.20, 0.50, 1.0, 2.5, 10 and 25 μg/ml using rat serum from a 10 mg/ml stock solution. Quality control samples (QCs) were used to evaluate the accuracy and precision during the qualification and were prepared in the same way as calibration standards at 0.10, 0.25, 2.0, and 10 μg/ml. The stable isotope labelled analogue DILLTQSPAILSVSPGER [13C615N4]  used as internal standard (IS) was added to calibration standard and QCs during sample preparation, after elution from the beads. All samples were submitted to the immunocapture and digestion steps prior to LC-MS/MS analysis.

Immunocapture and Digestion

For each sample 50 µl protein A magnetic beads was used. After removing the storage buffer and washing, 25 µl sample volume was loaded (60 min) on the beads in 200 µl PBS (phosphate buffered saline). A magnetic stand (Millipore) was used to separate the beads from the solution. After washing, proteins were eluted from the beads using 50 µl 100 mM glycine, pH 2.5. After removal of the beads, the pH was raised with TRIS (pH 8) and internal peptide solution was added (24 ng/ml end concentration). After reduction with dithiothreitol (50 mM end concentration) and alkylation with iodoacetamide (85 mM end concentration), trypsin (20 µg) was added in ammonium bicarbonate and the samples were incubated for 2.5 hours at 37°C. Finally, acetonitrile, formic acid and trifluoroacetic acid were added at final concentrations of respectively 10%, 1% and 0.5% to give an end volume of 205 µl.  Samples were transferred to an HPLC vial and injected (10 μl) into the chromatographic system.

Optimization of trypsin digestion

Digestion was carefully optimized since many factors, such as pH, ionic strength, temperature, and incubation time, can influence the efficiency of proteolysis. We chose TCPK-trypsin, which exhibits limited self-digestion and retains its activity over a long period. The optimal protein digestion was obtained at pH~8 in 100 mM ammonium bicarbonate buffer at a temperature of 37°C for 2.5 hours using 20 μg of trypsin. Protein denaturation was obtained by heating to 95°C, reduction was performed by addition of 50 mM dithiothreitol and incubation for 20 minutes at 56°C and alkylation was optimized by addition of 85 mM iodoacetamide and incubation for 30 minutes at ambient temperature in the dark.

Qualitative mass spectrometry for tryptic signature peptide identification

Peptide identification, for selection of the best candidates for the quantification of Infliximab, was performed using a Q-Exactive Orbitrap high resolution mass spectrometer (Thermo Fischer Scientific) coupled with a Proxeon EASY-nLC1000 pump (Thermo Fischer Scientific) or a Dionex Ultimate 3000 system (Thermo Fischer Scientific). Separation (Proxeon) was performed on a EasySPRAY column (Pepmap RSLC, 50 µm x 150 mm, 2 µm, Thermo Fisher Scientific) using a  reversed phase (10–40% acetonitrile, 0.1% formic acid) gradient (12 min.) running at 0.300 µL/min (Figure 1). Separation (Dionex) was performed on an Acquity HSS T3, column (100 x 2.1 mm, 1.8 µm, Waters) using a  reversed phase (2–30% acetonitrile, 0.1% formic acid) gradient (10 min) running at 0.500 mL/min (Figure 2). Full scan measurement was performed at a resolving power of 35000 with an AGC target of 3e6 across a scan range of 300–1100 m/z in profile mode, further settings: 2 µs scans,  max. injection time 200 ms mass tolerance 5 ppm. 

LC-MS quantitation of signature peptide

UPLC Chromatography was performed on an Acquity UPLC (Waters) using an Acquity HSS T3, 50 x 2.1 mm, 1.8 µm column (Waters). The column was eluted at a flow rate of 0.8 ml/min using as mobile phase A MilliQ water and as mobile phase B acetonitrile, both containing 0.1% (v/v) formic acid. The analytical gradient profile was as follows:

Slider image

[Click to enlarge]

The column temperature was 50°C, and the injection volume was 10 μl. The UPLC system was coupled to a XEVO TQS (Waters, Milford, MA) equipped with an electrospray ion source. Ionization was performed in the positive mode using a voltage of 600 V; the cone voltage was set to 30 V. Argon was used as the collision gas. SRM of DILLTQSPAILSVSPGER (DIL) and its internal standard DILLTQSPAILSVSPGER[13C615N4]  (DIL IS) was performed as follows: Peptide DIL 632.7 ([M+3H]3+ ions) => 545.3 (y5, collision energy 10 eV, quantifier) and 632.7 => 731.4 (y7, collision energy 18 eV, qualifier); Peptide DIL IS, 636.0 => 555.3 (y5, collision energy 10 eV, quantifier) and 636.0 => 741.4 (y7, collision energy 18 eV, qualifier). The dwell time for each SRM transition was 47 ms.  The quantification of Infliximab in serum was based on measurement of the DIL/DIL IS peak area ratio.

Quantification of Infliximab by the LC-MS/MS method in spiked rat serum

Calibration curves were established using linear regression and 1/x2 weighting, plotting the Infliximab concentration in the calibration standard samples (from 0.05 to 25 μg/ml) vs. the peak area ratio of the signature peptide from Infliximab and the signature peptide IS. Accuracy and precision of the method were evaluated during one day by analyzing the four QC samples (0.1, 0.25, 2.0, and 10 μg/ml) in five replicates. Assay accuracy (ratio of the measured concentration to the nominal concentration multiplied by 100) and precision (percent relative standard deviation, %RSD) were calculated at each concentration. The combined recovery of the immunocapture and digestion step  at each concentration was estimated from the signal of the signature peptide obtained after Infliximab extraction from rat serum followed by peptide release after tryptic digestion multiplied by molecular mass ratio of the signature peptide (multiplied by 2) and Infliximab and comparing this with the signal obtained for the added signature peptide IS.

Results and Discission

Selection of peptides for quantification

The strategy for quantification involved the selection of tryptic peptides from the variable regions of both the light and heavy chains. The corresponding sequences were submitted to similarity searches (Protein BLAST)[14] to ensure detection specificity. Table 1 lists all the potential peptides after in-silico digestion and BLAST searches vs. the species Rattus Norvegicus. 
The following peptide selection criteria were applied:
-Length between 6 and 20 amino acids
-No adjacent tryptic cleavage sites
-No methionine or cysteine in the peptide sequence
-No asparagine followed by glycine or serine
-Peptide contains part of variable domain

Figures and Tables

Slider image
Slider image
Slider image
Slider image
Slider image
Slider image
Slider image
Slider image
Slider image
Slider image

Figure 1. LC-MS Q Exactive full scan run of signature peptide DILLTQSPAILSVSPGER and its internal standard DILLTQSPAILSVSPGER[13C615N4] at 40 ng/ml (peptide level).

gfgffgfdgf

[Click to enlarge]

There are several other relevant peptide selection criteria, which were not found to be relevant for Infliximab and were thus not applied. Table 2 lists the results of the database search.The first practical step of the development was to identify specific peptides with the most intense MS signal, among those produced by digestion. These different peptides were monitored in an exploratory LC-MS run of an Infliximab digest using the Q-Exactive in the full scan mode. Peptide DILLTQSPAILSVSPGER exhibited a signal approximately three times higher than peptides ASQFVGSSIHWYQQR and TVAAPSVFIFPPSDEQLK. Finally, peptide DILLTQSPAILSVSPGER was selected for further experiments based on sensitivity and chromatography (e.g. peak shape), see Figure 2. A stable isotope labeled analogue of DILLTQSPAILSVSPGER was synthesized to serve as internal standard.

Immunocapture of Infliximab

We selected magnetic beads coated with protein A to pull down Infliximab via its human Fc part. The approximate absolute recovery observed after all the purification, reduction, alkylation and digestion steps was 30%, calculated using the theoretical internal standard concentration in the end extract and the theoretical concentration of the analyte peptide from Infliximab in the end extract, according to the equation below (Eq. 1)

R=((Sanal/SIS)/((Cprot*(MWpep*2/MWprot)/Dprot)/(CIS/DIS))) *100%          

Where R is the recovery in %, Cprot the concentration of protein in serum in ng/ml, MWpep the average molecular weight of the signature peptide in Da, MWprot the average molecular weight of the protein in Da, Dprot the protein dilution factor or the ratio between sample end volume and loading volume, CIS the concentration of the added IS solution in ng/ml, DIS the IS dilution factor or the ratio between end volume and added volume, Sanal the signal obtained for the analyte signature peptide and SIS the signal obtained for IS peptide.The results shown in this paper indicate that even with a relatively low absolute recovery, accurate results (with good relative recovery) can be obtained when quantifying proteins using calibration samples which have been prepared with the analyte protein in the matrix of interest and subjecting these to the same conditions as the study samples and QCs. A way to further optimize the repeatability of the method would be to load the protein A beads with stable isotope labeled protein internal standard instead of fortifying with peptide internal standard after elution of the proteins from the beads [16,17]. The parameter which requires further optimization in the presented method is the elution step from the protein A beads. When the protein A beads were subjected to another round of tryptic digestion a significant residual signal was observed for DILLTQSPAILSVSPGER (results not shown). In our experience, the elution step from the protein A beads was also critical in the method development for other monoclonal antibodies (results not shown). It is recommended to optimize these conditions instead of using the standard elution conditions described in protein A protocols.

Optimization of LC-MS/MS method for DILLTQSPAILSVSPGER

The most optimal transitions for DILLTQSPAILSVSPGER were selected on the basis of signal, background noise and chromatographic separation characteristics in the presence of matrix background. It should be noted that for DILLTQSPAILSVSPGER the most prominent charge state ([M+2H]2+ or [M+3H]3+) could change during method optimization. Especially, when a sensitive method is required, it is important to follow different precursor ion charge states during method optimization, also for other signature peptides. The results obtained with the calibration samples (prepared once, analyzed twice) are shown in Table 3 and the results of the QC samples are shown in Table 4.

The obtained calibration curve is presented in Figure 3.
Six blank serum samples of individual rats were analyzed, which exhibited a very low  background signal in the SRM transition of triply protonated DILLTQSPAILSVSPGER at QC LLOQ level.

In Figures 4-6 typical chromatograms are shown for blank rat serum, QC LLOQ and QC high samples, respectively. The obtained RSD for the QC low level was higher than 20%, although the RSD was lower than 20% (12%) when the results were calculated with the preset qualifier ion for DILLTQSPAILSVSPGER. 
Using the preset qualifier ion, however, the RSD for the LLOQ level was higher than 25% (28%). The presented method would therefore require further optimization prior to full validation according FDA or EMA guidelines.

Conclusion

This work presents an LC-MS/MS method for the quantification of the therapeutic monoclonal antibody Infliximab in rat serum. The approach is based on immunoprecipitation, tryptic digestion of Infliximab, and LC-MS/MS quantification of the signature peptide DILLTQSPAILSVSPGER and its stable isotopically labeled analogue IS. With state of the art UPLC and triple quadrupole instruments, a lower limit of quantification was of 100 ng/ml in rat serum was readily achievable, as illustrated in Figure 5. Since the peptide DILLTQSPAILSVSPGER originates from the variable region, it can also be utilized as a signature peptide for the quantification of Infliximab in higher order species, including Homo sapiens.

Acknowledgement

The authors would like to thank M. el Amrani from the University Medical Center in Utrecht for collaboration and for providing Infliximab solution.

References

1. Van Dongen WD, Niessen WM. LC-MS systems for quantitative bioanalysis. Bioanalysis 4(19), 2391-2399 (2012). [CrossRef]
2. Knutsson M, Schmidt R, Timmerman P. LC-MS/MS of large molecules in a regulated bioanalytical environment – which acceptance criteria to apply? Bioanalysis 5(18), 2211-2214 (2013).  [CrossRef]
3. Van Dongen WD, Renfurm LC. Conference report: moving forward together: “we are making progress”. Bioanalysis 6(9), 1159-1165 (2014). [CrossRef]
4. Hoofnagle AN, Wener MH. The fundamental flaws of immunoassays and potential solutions using tandem mass spectrometry. J Immunol Methods 347(1-2), 3-11 (2009). [CrossRef]
5. Van Den Broek I, Niessen WM, Van Dongen WD. Bioanalytical LC-MS/MS of protein-based biopharmaceuticals. J Chromatogr B Analyt Technol Biomed Life Sci 929, 161-179 (2013). [CrossRef]
6. Van Den Broek I, Blokland M, Nessen MA, Sterk S. Current trends in mass spectrometry of peptides and proteins: Application to veterinary and sports-doping control. Mass Spectrom Rev, (2013).[CrossRef] 
7. An B, Zhang M, Qu J. Toward sensitive and accurate analysis of antibody biotherapeutics by liquid chromatography coupled with mass spectrometry. Drug Metab Dispos 42(11), 1858-1866 (2014). [CrossRef]
8. Leary BA, Lawrence-Henderson R, Mallozzi C et al. Bioanalytical platform comparison using a generic human IgG PK assay format. J Immunol Methods 397(1-2), 28-36 (2013). [CrossRef]
9. Law WS, Genin JC, Miess C et al. Use of generic LC-MS/MS assays to characterize atypical PK profile of a biotherapeutic monoclonal antibody. Bioanalysis 6(23), 3225-3235 (2014). [CrossRef]
10. Zheng J, Mehl J, Zhu Y, Xin B, Olah T. Application and challenges in using LC-MS assays for absolute quantitative analysis of therapeutic proteins in drug discovery. Bioanalysis 6(6), 859-879 (2014). [CrossRef]
11. Jiang H, Zeng J, Titsch C et al. Fully validated LC-MS/MS assay for the simultaneous quantitation of coadministered therapeutic antibodies in cynomolgus monkey serum. Anal Chem 85(20), 9859-9867 (2013). [CrossRef]
12. Liu H, Manuilov AV, Chumsae C, Babineau ML, Tarcsa E. Quantitation of a recombinant monoclonal antibody in monkey serum by liquid chromatography-mass spectrometry. Anal Biochem 414(1), 147-153 (2011). [CrossRef]
13. Danese S, Colombel JF, Reinisch W, Rutgeerts PJ. Review article: infliximab for Crohn’s disease treatment–shifting therapeutic strategies after 10 years of clinical experience. Aliment Pharmacol Ther 33(8), 857-869 (2011). [CrossRef]
14. Altschul SF, Madden TL, Schaffer AA et al. Gapped BLAST and PSI-BLAST: a new generation of protein database search programs. Nucleic Acids Res 25(17), 3389-3402 (1997). [CrossRef]
15. Stadlmann JDI. Applications of a Mass-spectrometric Method for the N-glycosylation Analysis of Immunoglobulin G molecules (Doctoral Dissertation). Retrieved from https://zidapps.boku.ac.at/abstracts/download.php?dataset_id=7380&property_id=107. The University of Natural Resources and Applied Life Sciences, Vienna (2009).
16. Li H, Ortiz R, Tran L et al. General LC-MS/MS method approach to quantify therapeutic monoclonal antibodies using a common whole antibody internal standard with application to preclinical studies. Anal Chem 84(3), 1267-1273 (2012). [CrossRef]
17. Heudi O, Barteau S, Zimmer D et al. Towards absolute quantification of therapeutic monoclonal antibody in serum by LC-MS/MS using isotope-labeled antibody standard and protein cleavage isotope dilution mass spectrometry. Anal Chem 80(11), 4200-4207 (2008). [CrossRef]