OPEN-ACCESS PEER-REVIEWED

RESEARCH ARTICLE

Marieke G.G. Sturkenboom1, Henk van der Lijke1,2,‡,† , Erwin M. Jongedijk1,Wim Th. Kok2, Ben Greijdanus1, Donald R.A. Uges1, Jan-Willem C. Alffenaar1,†,*

1University of Groningen, University Medical Center Groningen, Department of Clinical Pharmacy and Pharmacology, Groningen, The Netherlands. 2University of Amsterdam, Van’t Hoff Institute for Molecular Sciences, Amsterdam, The Netherlands. * both first authors contributed equally. Present address: QPS Netherlands B.V., PO Box 137, 9700 AC Groningen, The Netherlands.

Journal of Applied Bioanalysis. Vol.1. No.3. pages 89-98 (2015)

Published 15 July 2015. https://doi.org/10.17145/jab.15.015 | (ISSN 2405-710X)

Download PDF

Correspondence:
*Alffenaar JW . University of Groningen, University Medical Center Groningen, Department of Clinical Pharmacy and Pharmacology, PO Box 30.001, 9700 RB Groningen, The Netherlands. Phone: +31 503614071, Fax: +31 503614087.

Citation:
Sturkenboom MGG, van der Lijke H, Jongedijk EM, KoK WTh, Greijdanus B, Uges DRA, Alffenaar JWC. Quantification of isoniazid, pyrazinamide and ethambutol in serum using liquid chromatography-tandem. J Appl Bioanal 1(3), 89-98 (2015).

Open-access and Copyright:
©2015 Sturkenboom MG et al. This article is an open access article distributed under the terms of the Creative Commons Attribution License (CC-BY) which permits any use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Funding/Manuscript writing assistance:
The authors have no financial support or funding to report and also declare that no writing assistance was utilized in the production of this article.

Competing interest:
The authors have declared that no competing interest exist.

Article history:
Received: 11 May 2015, Revised 15 June 2015, Accepted 16 June 2015

Abstract

Clinical studies on tuberculosis have shown a correlation between low drug exposure and treatment failure and acquired drug resistance. Objective was to develop a LC-MS/MS method for the quantification of isoniazid, pyrazinamide and ethambutol. Stable isotope-labelled isoniazid-D4 and ethambutol-D4 were used as internal standards. Protein binding of isoniazid, pyrazinamide and ethambutol was investigated and proved low. Therefore, sample preparation using ultrafiltration could be applied, resulting in linear calibration curves in the range of 0.2-8 mg/L for isoniazid and ethambutol and 2-80 mg/L for pyrazinamide. The method was validated according to the guidelines of the FDA.
A fast, simple and reliable LC-MS/MS method has been developed for the simultaneous determination of isoniazid, pyrazinamide and ethambutol in human serum for therapeutic drug monitoring and pharmacokinetic studies.

Keywords

Isoniazid, pyrazinamide, ethambutol, therapeutic drug monitoring, LC-MS/MS, tuberculosis 

Introduction

Tuberculosis (TB) is one of the infectious diseases with the highest morbidity and mortality in the world [1]. First-line treatment of TB consists of isoniazid, rifampin, pyrazinamide and ethambutol during the first two months, continuing with isoniazid and rifampin for another four months [1]. Although treatment is often successful, healthcare providers around the world are confronted with treatment failure daily. Recently, Pasipanodya et al. showed that risk on treatment failure was almost nine fold higher in patients with low drug exposure compared to patients with higher drug exposure [2]. Therefore, monitoring of drug exposure in patients at risk for low drug exposure is worthwhile to explore in a prospective manner [3].
To be able to evaluate drug exposure a fast, accurate and simple method for determination of isoniazid, rifampin, pyrazinamide and ethambutol in serum is valuable. Like for other antimicrobial drugs, it would be ideal to determine all four compounds in the same sample using a single method of analysis [4,5]. Song et al. published a method for the simultaneous determination of all first-line anti-TB drugs using liquid chromatography-tandem mass spectrometry (LC-MS/MS) [6]. Unfortunately, the polar compounds isoniazid, pyrazinamide and ethambutol elute near the void volume of the system, co-eluting with endogenous compounds resulting in substantial ion suppression [6,7]. As matrix effects may vary within and between subjects, ion suppression at the time of elution of the compounds of interest should be avoided or overcome using a stable isotope-labelled internal standard [8–10].
Zhou et al. used hydrophilic interaction chromatography (HILIC) to quantify isoniazid, rifampin, pyrazinamide, ethambutol and streptomycin [11]. They tested five different HILIC columns and several mobile phase compositions. An initial attempt by the authors to determine isoniazid and ethambutol using HILIC was abandoned, as our method suffered from split peaks and bad repeatability. Recently, Kim et al. developed a LC-MS/MS method for the simultaneous quantification of 20 anti-TB drugs. The method consisted of two different precipitation methods, different solid and mobile phases [12].
We successfully applied ultrafiltration as a means of sample preparation for the polar compound ribavirin [13] and we were interested whether this might also be applied for the first-line anti-TB drugs. However, rifampin shows high protein binding of around 80-90% [14], which makes it a less suitable candidate for ultrafiltration compared to isoniazid, pyrazinamide and ethambutol. Therefore, it was decided to split the patient sample, followed by a subsequent separate analysis of rifampin and a separate one for isoniazid, pyrazinamide and ethambutol together resulting in more reliable results within the same runtime. A previously developed method will be used for the analysis of rifampin [7,15]. The objective of this study was to develop and validate a simple and rapid LC-MS/MS method for the determination of isoniazid, pyrazinamide and ethambutol in human serum.

Experimental

Chemicals, reagents and disposables

Isoniazid (purity >99%) was purchased from Bufa (Uitgeest, The Netherlands). Ethambutol dihydrochloride (purity >99%) and pyrazinamide (purity ≥98%) were purchased from Sigma-Aldrich (Zwijndrecht, The Netherlands). The internal standard isoniazid-D4 was produced by CDN Isotopes (Pointe-Claire, Canada). The internal standard ethambutol-D4 was synthesized by Syncom (Groningen, the Netherlands). Acetonitrile (Ultra LC-MS-grade) was obtained from BioSolve (Valkenswaard, The Netherlands). Ammonium acetate and acetic acid (p.a. quality) from Merck were purchased from VWR (Amsterdam, The Netherlands). Water was purified with a Milli-Q system (Millipore Corporation, Billerica, MA, USA). Blank human serum and plasma was obtained from healthy volunteers according to the appropriate guidelines of the University Medical Center Groningen. Pall Nanosep 30K Omega centrifugal devices, Varian non-deactivated 2 mL borosilicate glass screw top vials and 600 μL polypropylene screw top vials were purchased from VWR (Amsterdam, The Netherlands). The inserts were made from Schott Fiolax®-clear glass and were purchased from Aluglas (Uithoorn, The Netherlands).

Method development

The analytical range was based on clinical pharmacokinetic data and set at 0.2–8 mg/L for isoniazid and ethambutol and 2–80 mg/L for pyrazinamide [3,16]. As stable isotope-labelled internal standards are known to compensate for matrix effects [9,10], isoniazid-D4 and ethambutol-D4 were used as such. Since a stable isotope-labelled internal standard for pyrazinamide was unavailable, no internal standard was used for the quantification of this compound.

Protein binding and ultrafiltration

Reported protein binding levels vary considerably for the three compounds: ranging from 0 to 74% for isoniazid [17,18] from 5 to 50% for pyrazinamide [17,19] and from 10 to 30% for ethambutol [17,20]. However, one expects protein binding to be limited as isoniazid, pyrazinamide and ethambutol are very polar compounds. If protein binding proves to be low, ultrafiltration as a means of deproteinization of the serum samples can be applied. Although ultrafiltration is commonly used for determining free unbound drug levels in serum [21], it has previously been applied as a means of sample preparation for other polar compounds such as acyclovir, ganciclovir, ribavirin and zidovudine [13,22–25]. The obtained ultra filtrate is purely aqueous which is particularly suitable for injecting onto LC-systems that are equilibrated under the same conditions. To investigate the applicability of ultrafiltration for the assay, it was decided to determine protein binding of the three compounds using the 30 kDa centrifuge filters applied for the sample preparation. The Pall Nanosep 30 K Omega centrifugal device was chosen because it removes over 99.5% of all serum proteins, shows low protein binding, allows for fast processing times and fits standard micro-centrifuges [13].
Protein binding was determined by measuring a calibration curve with calibration standards at seven levels prepared in human serum and compare it to an equivalent calibration curve prepared in water with added blank ultra filtrate to compensate for matrix effects. The ratio of the slopes of these curves shows the extent of protein binding. In this experiment, calibration curve samples were freshly prepared and processed within 30 minutes to minimise alterations to isoniazid other than the reversible protein binding under investigation. This period is comparable to other studies on protein binding [26,27]. Inert type 1 borosilicate glass inserts were used to minimise any adsorption of ethambutol to glass inserts.

Adsorption

During early method development, a significant decline in signal for ethambutol was observed when the aqueous test mixtures were injected from the standard non-deactivated glass vials compared to plastic vials. To investigate this observation, an adsorption test was performed. A test solution with isoniazid, pyrazinamide and ethambutol was prepared and was divided over a polypropylene vial, a glass vial and a glass insert made out of type I borosilicate glass. The samples were shaken vigorously for one minute and five repetitive injections were made from each type of vial.

Carry-over

Carry-over from high isoniazid and ethambutol standards was observed during method development. Therefore, an experiment in fivefold was performed to evaluate the magnitude of the carry-over by injecting five LLOQ (Lower Limit of Quantification) samples after a HLQ (Higher Limit of Quantification with 8.0 mg/L isoniazid and ethambutol and 80 mg/L pyrazinamide) sample.

Sample preparation

For the calibration (reference) standards and Quality Control (QC) samples, separate stock solutions of 200 mg/L isoniazid and ethambutol and 2000 mg/L pyrazinamide in water were prepared. Calibration standards and QC samples were freshly prepared immediately before use by diluting appropriate amounts of stock solutions with blank human serum or plasma. QC samples were freshly prepared as stability of isoniazid is known to be limited [28,29]. The total amount of stock solutions added to the serum was less than 5% of the final volume. Calibration standards at eight levels ranged from 0.2-8.0 mg/L for isoniazid and ethambutol and from 2.0-80 mg/L for pyrazinamide. QC samples consisted of LLOQ, Low, Medium, High and over-the-curve (OTC) levels for each compound (Table 3). The internal standard solution was prepared with 0.25 mg/L isoniazid-D4 and 0.25 mg/L ethambutol-D4 in water. QC samples for the freeze/thaw stability were prepared beforehand. Separate Low and High level QC samples in human serum were thawed, frozen and analysed on three consecutive days. The samples for determining the bench top stability were freshly prepared and analysed immediately and after being kept for 24h at ambient temperature. Calibration standards, QC or human serum samples were homogenised and aliquots of 10 µL were directly transferred onto the upper reservoir of the centrifuge filters and 250 µL of internal standard solution was added. The centrifuge filters were closed and the samples were briefly homogenised using a vortex mixer. Filtration was done by centrifugation for 10 minutes at 14000 g at an angle of 45º. A 5 µL volume of the ultra filtrate was injected onto the LC-MS/MS system. Samples for the stability test were processed in the same way as the samples for the validation and routine analysis.

Analytical validation

In accordance with the ‘Guidance for Industry – Bioanalytical Method Validation’ of the FDA [30], method validation included selectivity, linearity, accuracy, precision and stability. The linearity was assessed by using eight point calibration curves on three days for the three compounds. To determine accuracy and precision, LLOQ, Low, Medium and High QC samples were analysed. All QC samples were analysed in fivefold in three separate runs on separate days. The calibration model was chosen by evaluating the accuracy results at LLOQ, Low, Medium and High concentration level of the first precision and accuracy run. The simplest linear model at three weighting factors (x, 1/x and 1/x2) at which all accuracies were within the specifications of the FDA was chosen and set for the entire validation and ensuing bioanalysis. Automatic immuno-assays do not exist for the quantification of anti-TB drugs. Therefore, all assays for anti-TB drugs are ‘in-house’ developed and validated chromatographic methods [31]. To be able to compare obtained results between laboratories, it is important to participate in an external quality control programme. Recently, such a proficiency testing programme has been set up [31] and we participated in the programme with the developed method.

Matrix effect

The matrix effect was assessed using six different human sera and six different human edetate plasmas adding up to 12 different matrices. The first set (A) was prepared of QC stock in water to evaluate the neat MS response of the compounds and the internal standards. The second set (B), 12 different calibration sets were prepared with internal standard solution and after ultrafiltration QC stock was added. The matrix factor was calculated as the ratio of the peak area of the spiked standards after sample preparation (B) to the corresponding peak area of the neat solution standards (A) [9]. The internal standard-normalised matrix factor was calculated by dividing matrix factor of isoniazid and ethambutol by the matrix factor of their internal standard [9]. As pyrazinamide was determined without internal standard, the matrix factor was calculated without normalisation.

Stability

The stability of the compounds was extensively evaluated as a part of a larger investigation on the stability of many anti-TB drugs. Human serum standards were prepared by adding appropriate amounts of stock solution to blank matrices. Analyses were performed in duplicate and time point 0 served as the reference. The final serum concentrations for the Low concentration samples were 0.51 mg/L isoniazid, 20.2 mg/L pyrazinamide and 1.00 mg/L ethambutol. For the High concentration samples, the values were 6.89 mg/L, 62.4 mg/L and 5.09 mg/L for isoniazid, pyrazinamide and ethambutol respectively.The prepared human serum standards were divided into 0.5 mL aliquots and stored in sealed plastic tubes at four different conditions: ambient temperature (18-24°C), refrigerator (4-8°C) and at two freezing temperatures (-20°C and -80°C). Frozen samples were thawed immediately before analysis. To determine bench top stability, samples were thawed and analysed immediately and after being left for 24 hours at ambient temperature. To determine the autosampler stability, Low and High QC samples were processed at day one, were left in the autosampler for 24 hours and re-injected on day two. To determine freeze/thaw stability, separate Low and High QC samples were thawed, frozen at -80°C and analysed on three consecutive days. One portion of human serum was divided into 1 mL aliquots in 10 mL glass containers and freeze-dried using a Leybold Lyovac GT4 lyophilizer according to local standard operating procedure by authorized personnel. Then they were sealed and refrigerated at 4-8°C. Freeze-dried samples were reconstituted by adding 1 mL of water, immediately before use.

Clinical application

After validation the analytical procedure was used in routine patient care and a clinical study evaluating the influence of food and fasting on the pharmacokinetics of first-line anti-TB drugs. Twenty patients included in this study were diagnosed with drug susceptible TB and provided written informed consent. The study protocol was approved by the Medical and Health Research Ethics Committee of the Gadjah Mada University in Yogyakarta, Indonesia.

Results

Method development

Using the Waters Atlantis T3 column, best peak shapes and repeatable retention times were obtained by starting and ending the mobile phase gradient in 100% aqueous conditions. Mean retention times were 1.15 min for ethambutol, 1.41 min for isoniazid and 1.72 min for pyrazinamide, respectively (Table 1). The hold-up time of the system is approximately 0.9 min. To optimize the chromatography of the compounds a mobile phase buffer pH 5 was used. This resulted in acceptable peak shapes  (Fig. 1). Although isoniazid and ethambutol elute within a short period, no interference was observed. However, some tailing of ethambutol was observed. We were unable to further diminish this tailing. The results of the protein binding tests are presented in  Table 4. As protein binding indeed proved to be low, ultrafiltration as a means of sample preparation could be applied. In the adsorption test, ethambutol showed a significant loss of approximately 35% in peak area when injected from the standard glass vial  (Fig. 2). The isoniazid response showed a small but significant decrease when injected from the polypropylene and the glass vial. Pyrazinamide was unaffected by the different materials. The type I borosilicate glass insert showed the least absorption for most compounds and was chosen as the standard vial. The test for carry-over showed that it was caused by a memory effect of the column. This memory effect was proven by repeating the gradient three times during an analytical run. During this experiment, isoniazid and ethambutol peaks were observed at the interval of the original gradient. At every interval, peak areas were approximately 10% of the preceding peak, for both isoniazid and ethambutol. This situation was similar to the carry-over our group observed during the quantification of rifampin [15]. The memory effect was minimized to approximately 0.1% by adding two flushing and conditioning steps to the original gradient, in which the column is flushed with 100% eluent B and conditioned with 100% eluent A.

Figures and Tables

Slider image
Slider image
Slider image
Slider image
Slider image
Slider image
Slider image
Slider image
Slider image
Slider image

[Click to enlarge]

Analytical validation

Regarding selectivity, examination of the six independent blank human serum or plasma samples revealed no interfering components at the mass transitions monitored  (Fig. 1).
The validation results regarding accuracy and precision are listed in  Table 5. Calibration curves for isoniazid and ethambutol were obtained by plotting the ratio of peak area to that of the internal standard against the concentration. Pyrazinamide was not analysed using an internal standard because no stable isotope-labelled pyrazinamide was available. The calibration curves were fitted using least squares linear regression with a weighting factor of 1/x. None of the calibrators was discarded in the validation procedure. The calibration curves were linear in the range of 0.2-8 mg/L for isoniazid and ethambutol and 2-80 mg/L for pyrazinamide. The average correlation coefficients were 0.9987 for isoniazid, 0.9992 for pyrazinamide and 0.9984 for ethambutol. All the calibration curves showed a significant goodness of fit. All obtained results met the acceptance criteria set by the FDA [27]. These criteria were < 20% bias and <20% coefficient of variation (CV) for LLOQ QC samples and <15% bias and <15% CV for the other QC samples. The maximum overall bias was 10.7% and the maximum overall CV was 8.1%, which is well within the limits set by the FDA. We participated with this method in the second round of an interlaboratory quality control program for the measurement of anti-TB drugs [31]. All values were within the limits of 20% set by the organisation of the programme.

Matrix effect

The results for the matrix effect tests are listed in Table 6. Data for both serum and plasma samples were pooled, as no significant difference was observed between them. For all compounds, a small positive (internal standard-normalised) matrix factor was determined for low and high concentrations. Maximum CV was 6.0% for the Low level of ethambutol, this is within the limit of 15% set by the EMA [9].

Stability

The results of the stability tests showed that isoniazid is not stable at ambient temperature in human serum  (Table 7). However, if it is present in ultra filtrate as in the autosampler test, isoniazid proves to be stable for at least 24 h. Isoniazid serum standards are only stable at -80°C or in freeze-dried conditions for longer periods. From the data in Table 7, it is calculated that average in vitro half-life values for isoniazid were 6 days at ambient temperature, 15 days at refrigerating temperatures and 115 days at -20°C. None of the three compounds in human serum showed degradation beyond criteria after three freezing and thawing cycles, although the concentrations of the isoniazid and pyrazinamide samples at low concentration level dropped considerably after the third cycle. Pyrazinamide in human serum is stable at all temperatures except at ambient temperature. Measured pyrazinamide concentrations at ambient temperature started to decline significantly after five days. Since the human matrices stored at ambient temperature showed signs of decomposition over time, degradation products may have affected these measurements. None of the measured ethambutol concentrations fell beyond acceptance limits at any of the time points for all storage conditions. A limited concentration drop for all long term stability samples was observed however.

Clinical application

To illustrate the clinical application a concentration-time curve over a 24 h time period of a TB patient recently started on TB treatment is shown in  Fig. 3. She received her medication orally on an empty stomach. Including the preparation of calibration standards and QC samples, it takes approximately 1.5 h to analyse the three drugs in these 12 samples. The maximum capacity is almost 200 samples per 24 h. In the same time, the samples can be analysed for rifampin on a separate LC-MS/MS [7,15] resulting in a processing time similar to the method of Song [6].

Discussion

The most important aspect from these results presented here is that ethambutol did not show any significant in vitro protein binding in serum using this method, which clearly contradicts earlier results [17,20]. This may partly be caused by the difference in temperature, as this factor is known to influence protein binding [26]. In our experiments we used an unthermostated centrifuge. However it can be observed from the accuracy and precision results in  Table 5 that temperature had limited influence on the validation. In the adsorption test, ethambutol showed a significant loss of approximately 35% in peak area when injected from the standard glass vial. Earlier reported protein binding results [17,20] vary and may be obscured by adsorption of ethambutol to non deactivated glass vials. Furthermore, any possible protein binding would be reduced by the extensive dilution of the samples with internal standard solution during sample preparation before the ultrafiltration [32]. By then, variation of any possible protein binding of ethambutol in the diluted samples is compensated by the stable isotope-labelled internal standard. Protein binding of isoniazid and pyrazinamide is also lower than reported values. Reported (reversible) protein binding values of isoniazid might have been obscured by the instability of isoniazid in serum. The instability of isoniazid is attributed to irreversible protein binding [33]. As protein binding of isoniazid, pyrazinamide and ethambutol proved low, ultrafiltration as a means of sample preparation could be applied resulting in a fast and simple method of analysis. Significant carry-over and memory effects were observed for isoniazid and ethambutol, during method development. This can be explained by absorption to silica and metal surfaces, due to their polar characteristics. Carry-over was overcome by rinsing the sample needle and needle seat with eluent B in-between sample injections. Memory effects were overcome by adding two flushing and conditioning steps to the original gradient (Table 2). Bench-top stability testing and the more extensive stability assessment showed that isoniazid is unstable in human serum unless it is freeze-dried or stored at -80°C. Instability of isoniazid in plasma at ambient temperature and -20ºC has been reported earlier [28,29]. This instability calls for adequate standard operating procedures and rapid handling of samples. The autosampler stability testing show that isoniazid is stable in ultra filtrate for at least 24 hours proving that serum proteins are involved in the instability of isoniazid. Pyrazinamide serum samples are more stable and can be kept for at least 51 days in a refrigerator. Ethambutol proved to be stable at ambient temperature. Using this method, almost 200 patient samples can be analysed for isoniazid, pyrazinamide and ethambutol in 24 h. If one makes use of an optimal sampling strategy consisting of three samples to calculate the area under the concentration-time curve, information on drug exposure of 65 patients can be obtained in one day. The presented analytical method has shown to be robust for therapeutic drug monitoring and pharmacokinetic studies of isoniazid, pyrazinamide and ethambutol. It may be of clinical help and importance as the current opinion is moving from standard dosing towards a more drug exposure based evaluation of TB treatment [34].

Conclusion

A fast, simple and reliable LC-MS/MS method has been developed for the determination of isoniazid, pyrazinamide and ethambutol in human serum. Isoniazid, pyrazinamide and especially ethambutol have proven to be low protein bound. Therefore, ultrafiltration as a means of sample preparation could be applied. The method adheres to all the validation criteria and is suitable for the simultaneous determination of isoniazid, pyrazinamide and ethambutol in human serum for therapeutic drug monitoring and pharmacokinetic studies.

References

1.World Health Organization STBP. The global plan to stop TB 2011-2015: transforming the fight towards elimination of tuberculosis. http://www.stoptb.org/assets/documents/global/plan/tb_globalplantostoptb2011-2015.pdf.
2. Pasipanodya JG, McIlleron H, Burger A, Wash PA, Smith P, Gumbo T. Serum drug concentrations predictive of pulmonary tuberculosis outcomes. J Infect Dis 208(9), 1464–1473 (2013). [CrossRef]
3. Peloquin CA. Therapeutic drug monitoring in the treatment of tuberculosis. Drugs 62(15), 2169–2183 (2002). [CrossRef]
4. Alffenaar JW, Wessels AM, van Hateren K, Greijdanus B, Kosterink JG, Uges DR. Method for therapeutic drug monitoring of azole antifungal drugs in human serum using LC/MS/MS. J Chromatogr Anal Technol Biomed life Sci 878(1), 39–44 (2010). [CrossRef]
5. Mistri HN, Jangid AG, Pudage A, Gomes N, Sanyal M, Shrivastav P. High throughput LC-MS/MS method for simultaneous quantification of lamivudine, stavudine and nevirapine in human plasma. J Chromatogr Anal Technol Biomed life Sci 853(1-2), 320–332 (2007). [CrossRef]
6. Song SH, Jun SH, Park KU, et al. Simultaneous determination of first-line anti-tuberculosis drugs and their major metabolic ratios by liquid chromatography/tandem mass spectrometry. Rapid Commun Mass Spectrom 21(7), 1331–1338 (2007). [CrossRef]
7. De Velde F, Alffenaar JW, Wessels AM, Greijdanus B, Uges DR. Simultaneous determination of clarithromycin, rifampicin and their main metabolites in human plasma by liquid chromatography-tandem mass spectrometry. J Chromatogr Anal Technol Biomed life Sci 877(18-19), 1771–1777 (2009). [CrossRef]
8. Matuszewski BK, Constanzer ML, Chavez-Eng CM. Strategies for the assessment of matrix effect in quantitative bioanalytical methods based on HPLC-MS/MS. Anal Chem 75(13), 3019–3030 (2003). [CrossRef]
9. European Medicines Agency C for MP for HU. Guideline on bioanalytical validation (EMEA/CHMP/EWP/192217/2009). European Medicines Agency, London. http://www.ema.europa.eu/docs/en_GB/document_library/Scientific_guideline/2011/08/WC500109686.pdf
10. Matuszewski BK. Standard line slopes as a measure of a relative matrix effect in quantitative HPLC-MS bioanalysis. J Chromatogr 830(2), 293–300 (2006). [CrossRef]
11. Zhou Z, Wu X, Wei Q, et al. Development and validation of a hydrophilic interaction liquid chromatography-tandem mass spectrometry method for the simultaneous determination of five first-line antituberculosis drugs in plasma. Anal Bioanal Chem 405(19), 6323–6335 (2013). [CrossRef]
12. Kim HJ, Seo KA, Kim HM, et al. Simple and accurate quantitative analysis of 20 anti-tuberculosis drugs in human plasma using liquid chromatography-electrospray ionization-tandem mass spectrometry. J Pharm Biomed Anal 102C, 9–16 (2014). [CrossRef]
13. Van der Lijke H, Alffenaar JW, Kok WT, Greijdanus B, Uges DR. Determination of ribavirin in human serum using liquid chromatography tandem mass spectrometry. Talanta 88, 385–90 (2012). [CrossRef]
14. Becker C, Dressman JB, Junginger HE, et al. Biowaiver monographs for immediate release solid oral dosage forms: rifampicin. J Pharm Sci 98(7), 2252–2267 (2009). [CrossRef]
15. Vu DH, Koster RA, Wessels AM, Greijdanus B, Alffenaar JW, Uges DR. Troubleshooting carry-over of LC-MS/MS method for rifampicin, clarithromycin and metabolites in human plasma. J Chromatogr Anal Technol Biomed life Sci 917-918, 1–4 (2013). [CrossRef]
16. Chideya S, Winston CA, Peloquin CA, et al. Isoniazid, rifampin, ethambutol, and pyrazinamide pharmacokinetics and treatment outcomes among a predominantly HIV-infected cohort of adults with tuberculosis from Botswana. Clin Infect Dis 48(12), 1685–1694 (2009). [CrossRef]
17. Micromedex® 2.0, (electronic version). Thomson Reuters (Healthcare), Greenwood Village, Colorado, USA. Available at: http://www.thomsonhc.com
18. Becker C, Dressman JB, Amidon GL, et al. Biowaiver monographs for immediate release solid oral dosage forms: isoniazid. J Pharm Sci 96(3), 522–531 (2007). [CrossRef]
19. Becker C, Dressman JB, Amidon GL, et al. Biowaiver monographs for immediate release solid oral dosage forms: pyrazinamide. J Pharm Sci 97(9), 3709–3720 (2008). [CrossRef]
20. Becker C, Dressman JB, Amidon GL, et al. Biowaiver monographs for immediate release solid oral dosage forms: ethambutol dihydrochloride. J Pharm Sci 97(4), 1350–1360 (2008). [CrossRef]
21. Singh SS, Mehta J. Measurement of drug-protein binding by immobilized human serum albumin-HPLC and comparison with ultrafiltration. J Chromatogr Anal Technol Biomed life Sci 834(1-2), 108–116 (2006). [CrossRef]
22. Nebinger P, Koel M. Determination of acyclovir by ultrafiltration and high-performance liquid chromatography. J Chromatogr 619(2), 342–344 (1993). [CrossRef]
23. Koel M, Nebinger P. HPLC determination of serum ganciclovir using ultrafiltration, ultraviolet and fluorescence detection. J Pharm Biomed Anal 12(3), 429–432 (1994). [CrossRef]
24. Nebinger P, Koel M. Determination of serum zidovudine by ultrafiltration and high-performance liquid chromatography. J Pharm Biomed Anal 12(1), 141–143 (1994). [CrossRef]
25. Kenney KB, Wring SA, Carr RM, Wells GN, Dunn JA. Simultaneous determination of zidovudine and lamivudine in human serum using HPLC with tandem mass spectrometry. J Pharm Biomed Anal 22(6), 967–983 (2000). [CrossRef]
26. Stove V, Coene L, Carlier M, De Waele JJ, Fiers T, Verstraete AG. Measuring unbound versus total vancomycin concentrations in serum and plasma: methodological issues and relevance. Ther Drug Monit 37(2), 180–7 (2015). [CrossRef]
27. Butterfield JM, Patel N, Pai MP, Rosano TG, Drusano GL, Lodise TP. Refining vancomycin protein binding estimates: identification of clinical factors that influence protein binding. Antimicrob Agents Chemother 55(9), 4277–82 (2011). [CrossRef]
28. Huang L, Marzan F, Jayewardene AL, Lizak PS, Li X, Aweeka FT. Development and validation of a hydrophilic interaction liquid chromatography-tandem mass spectrometry method for determination of isoniazid in human plasma. J Chromatogr Anal Technol Biomed life Sci 877(3), 285–290 (2009). [CrossRef]
29. Hutchings A, Monie RD, Spragg B, Routledge PA. A method to prevent the loss of isoniazid and acetylisoniazid in human plasma. Br J Clin Pharmacol 15(2), 263–266 (1983). [CrossRef]
30. Food and Drug Administration, U.S. Department of Health. Guidance for industry, bioanalytical method validation (2001). http://www.fda.gov/downloads/Drugs/Guidances/ucm070107.pdf
31. Aarnoutse RE, Sturkenboom MG, Robijns K, et al. An interlaboratory quality control programme for the measurement of tuberculosis drugs. Eur Respir J. 2015 Apr 16. [Epub ahead of print]. http://dx.doi.org/10.1183/09031936.00177014
32. Yuan J, Yang DC, Birkmeier J, Stolzenbach J. Determination of protein binding by in vitro charcoal adsorption. J Pharmacokinet Biopharm 23(1), 41–55 (1995). [CrossRef]
33. Holdiness MR. Chromatographic analysis of antituberculosis drugs in biological samples. J Chromatogr 340, 321–359 (1985). [CrossRef]
34. Sotgiu G, Alffenaar J-W, Centis R, et al. Therapeutic drug monitoring: how to improve drug dosage and patient safety in tuberculosis treatment. Int J Infect Dis 32, 101–4 (2015). [CrossRef]